The Ono Sokki TS-3200A is greatly improved in setting and usability with large sized LCD which offers easy and clear view of good contrast. Setting conditions and measurement data can be displayed in the same screen.
Ten torque detection settings can be stored in memory, which is very convenient function when you use several torque detectors. The detector setting can be changed easily at the time for replacement.
Items on a main display can be selected and displayed simultaneously up to 7 items. As well as torque, rotation and output (W) of calculated data, each of maximum value, minimum value, peak to peak value, and ripple ratio can be selected to display. For example, average value and fluctuation range of torque can be monitored at the same time. In addition, the TS-3200A provides a variety of interfaces such as data outputs of 2 channels with analog voltage as standard, digital interfaces and additional functions as options. It may expand possibility of data communication with external equipment for control and data processing.
Features
Overview specifications
Amplification section |
Isolated unbalanced DC amplification |
---|---|
Input signal |
Phase-differential type torque detector output signal |
Input impedance |
Approx. 2 kΩ (when input signal amplitude at 2 Vpp or less) |
Input frequency range |
200 Hz to 50 kHz |
Input signal amplitude range |
0.2 Vpp to 15 Vpp |
Connector |
TRC116 - 23A10 - 7F |
Capacity |
1 to 9999 |
Factor |
1 to 65535 |
Unit |
mNm, Nm, kNm |
Number of detection gears |
1 to 9999 P/R |
N-0 compensation |
Stored in each memory for CW and CCW directions |
Response |
Time constant can be settable among; 16 ms/31 ms/63 ms/125 ms/250 ms/500 ms/1 s/2 s/4 s/8 s/16 s/32 s/64 s |
Number of display digits |
Polarity + 4-digit or polarity + 5-digit, selectable |
Measurement clock |
4.195 MHz |
Sample clock |
Sampling time for 1 data: 1/256 Hz (approx. 4 ms) |
Number of display digits |
Polarity + 4-digit or 5-digit, selectable |
Unit |
mNm, Nm, kNm |
Display gate time setting |
1 s to 10 s (shared with a rotating section)or external gate |
Accuracy |
When combined with the detector (with 1 s gate time)
|
Analog output |
Voltage output 0 to ±10 V/F.S.
|
Response |
16 ms to 64 s depending on the time constant setting |
Accuracy |
When used in combination with a torque detector (The display renewal rate is 1 s.)
|
Temperature drift |
±0.01 %/F.S./ °C |